Skip to main content
Recall Canary
TD Navio Midi Speech Generating Device
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

TD Navio Midi Speech Generating Device

Tobii Dynavox recalls TD Navio Midi speech device over battery tab damage risk.

What to do

Contact Tobii Dynavox for instructions on returning or servicing your device. Stop using the device if you notice any signs of overheating or damage.

Recalled
Listed
Updated
Recall No.
Z-2959-2026
Agency
FDA
Firm
Tobii Dynavox Llc
Recall ID
99410
Status
ONGOING

Why it was recalled

Tobii Dynavox is recalling the TD Navio Midi Speech Generating Device, model/catalog number 1000798, because the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure and potentially cause an internal short circuit, localized heating, and gas venting.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

Lot code: 20260615

ProductRecall #UDI-DI
TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799Z-2960-2026240904000016, 260617070001, 260617070001, 20260615
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798Z-2959-2026240628000004, 260617070001, 260617070001, 20260615

Where it was sold

Distributed nationwide.

Timeline

  1. Firm initiated recall
  2. Publicly announced
  3. Listed on Recall Canary
  4. Summary updated

Other Device Defect recalls

← Back to all recalls

Editorial summary by Recall Canary · summary updated 2026-08-16 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification