Skip to main content

Ongoing FDAEvent 99630 (29 days ago)

Boston Scientific StoneSmart Connect Console

The StoneSmart Connect Console has power adapters with higher-than-expected resistance in the electrical grounding path.

Moderate risk

Using it may cause temporary or medically reversible health problems. Serious harm is unlikely.

FDA Class II recall · How risk is rated

What to do

What happened

Boston Scientific Corporation, based in Marlborough, Massachusetts, voluntarily recalled its StoneSmart Connect Console on July 22, 2026. The recall was initiated because a subset of power adapters used with the console may exhibit higher-than-expected resistance in the electrical grounding path, which could cause the power adaptor to fail electrical safety testing during installation. The recall involves 982 units and was added to the FDA's Enforcement Report on September 16, 2026.

The affected consoles were distributed worldwide, including nationwide in the United States across states such as Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, West Virginia and Wyoming.

International distribution included Australia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Lebanon, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkiye, Canada, Hong Kong, Singapore, Taiwan, Argentina, Chile, and Japan. Customers were notified by letter.

The FDA classified the recall on September 16, 2026, and it was in the weekly enforcement report on September 23, 2026.

Summary by Recall Canary from the FDA notice.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Compare the details below with your product and check the FDA notice to confirm whether it is affected.

  1. Product details

    Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DVI to HDMI Cable, (1) DisplayPort Cable REF: M0067933000, M006793300H0, M006793300R0, M006793300D0, M006793300Z0, M0067930000, M006793000R0

    UDI-DI
    • 00191506002510
    • 00191506021771
    • 00191506002558
    • 00191506002534
    • 00191506021788
    • 00191506002527
    • 00191506002541
    • 00191506022549
    Lot
    All Serial Numbers: REF: M0067933000 (UDI 00191506002510, UDI (00191506021771)

Details

Company began the recall
Sold in

Worldwide - US Nationwide distribution including in the states of Arizona, Florida, California, Oregon, Illinois, North Carolina, Nebraska, Utah, Colorado, Ohio, Arkansas, New Mexico, Alabama, Wisconsin, Tennessee, Indiana, Georgia, Ohio, Wisconsin, Kentucky, Connecticut, Kansas, South Carolina, Illinois, Nebraska, Missouri, Mississippi, Delaware, Texas, New York, West Virginia, Massachusetts, Montana, Washington, Wyoming, Nevada, Hawaii, Idaho, Missouri, Maine, West Virginia, Nebraska, Oklahoma, Minnesota, Louisiana, West Virginia, Nevada, Pennsylvania, New Hampshire, New Jersey, Michigan, Missouri, Alaska, Virginia, Mississippi, Wyoming, Maine and the countries of Australia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Lebanon, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkiye, Canada, Hong Kong, Singapore, Taiwan, Argentina, Chile, Japan.

Event ID
99630
Recall No.
Z-3188-2026
Company
Boston Scientific CorporationMarlborough, Massachusetts
Quantity recalled
982 units
Recall type
Voluntary: Firm initiated
Customers told by
Letter
  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. Listed on Recall Canary

  4. In FDA's weekly enforcement report

  5. Summary updated

All recalls from Boston Scientific Corporation

Summary by Recall Canary, written with AI · summary updated Oct 3, 2026 · primary source: U.S. Food and Drug Administration