Canon Alphenix Ceiling Movement Gear Screws
Canon recalls Alphenix X-ray systems (INFX-8000C/F/V) over loose ceiling gear screws.
Contact Canon Medical Systems to schedule inspection and repair of the ceiling movement gear screws. Do not use the system until the screws are verified and tightened.
Recall at a glance
- Quantity recalled
- 10 systems
Why it was recalled
Canon Medical Systems is recalling Alphenix X-ray interventional systems because the fixing screws of the ceiling movement gear in the ceiling-suspended arm may become loose, potentially causing the arm to move unexpectedly. The affected models are INFX-8000C, INFX-8000F, and INFX-8000V.
Is mine affected?
Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.
Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device
- UDI-DI
- 30101165, 30093711, 30029686, 79000002, 30049491, 30100053, 30059175, 30059180, 30095017, 30097052
Where it was sold
Distributed nationwide.
Timeline
- Firm initiated recall
- Publicly announced
- Listed on Recall Canary
- Summary updated
Official source
Other Device Defect recalls
Editorial summary by Recall Canary · summary updated 2026-08-26 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification