FDA recall, , Moderate risk
Fisher Diagnostics Infinity Ammonia Reagent and Standard Set
Infinity Ammonia Reagent and Standard Set have an ammonia reference standard with an incorrect concentration, causing a low bias.
What to do
Check yoursIf a code on your device is listed below:
- Check the lot against the recall notice and ask your provider or supplier before you stop using it.
Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.
What happened
Fisher Diagnostics of Middletown, Virginia, is recalling its Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers and its Infinity Ammonia Standard Set. The company started the recall on its own. An ammonia reference standard used to make the reagent was assigned an incorrect concentration.
This causes a low bias of about 6% to 15%. The agency states this could lead to incorrect patient classification, delays in treatment, and errors in clinical treatment decisions, which may cause patient harm.
The reagent and standard set are used in laboratory tests. The products were distributed nationwide in the United States, as well as worldwide, including Puerto Rico. The company began the recall on October 21, 2026.
Is your set affected?
2 products
Match the UDI-DI and lot code listed under your product.
A UDI-DI is the device identifier printed with the barcode on the device label.
Infinity Ammonia Standard Set
Product details from the notice
Infinity Ammonia Standard Set Model/Catalog Number: TR60901 Product Description: In Vitro Diagnostic calibrator and control
- UDI-DI
- 00845275003314
- Lot code
- 382963
- 364987
- 307166
Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers
Product details from the notice
Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers Model/Catalog Number: OSR61154 Product Description: In Vitro Diagnostic reagent, calibrator, and control
- UDI-DI
- 00845275000016
- Lot code
- 2915
- 2916
- 9511
- 9510
- 0644
- 0645
- 0614
- 0613
- 5301
- 9901
- 2525
- 7602
- 7603
- 4111
- 2393
- 7411
- 6624
- 3223
- 3231
- 7779
- 7780
- 0206
- 0209
Recall details
- Posted by FDA
- Recall began
- Status
- Open
- Weekly archive
- Listed the week of Nov 23, 2026
- Sold in
Worldwide - US Nationwide distribution in the states and territory of California, Illinois, Pennsylvania, Puerto Rico and the countries of Australia, GER, South Korea.
- Recalled by
- Fisher DiagnosticsMiddletown, Virginia
- Hazard
- Device defect
- Risk
Moderate risk
Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.
FDA Class II recall · How risk is rated
- FDA event number
- 100164
- Recall numbers
- Z-0533-2027 to Z-0534-2027
- Recall type
- Started by the company
- How the company first told customers
- More than one way: email, fax, letter, press release, phone or visit
All recalls from Fisher Diagnostics
Timeline
FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.
Fisher Diagnostics began the recall
FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)
Added to Recall Canary