Fisher Diagnostics Infinity Ammonia Reagent and Standard Set

Infinity Ammonia Reagent and Standard Set have an ammonia reference standard with an incorrect concentration, causing a low bias.

Product photo

What to do

Check yours

If a code on your device is listed below:

  1. Check the lot against the recall notice and ask your provider or supplier before you stop using it.

Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.

What happened

Fisher Diagnostics of Middletown, Virginia, is recalling its Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers and its Infinity Ammonia Standard Set. The company started the recall on its own. An ammonia reference standard used to make the reagent was assigned an incorrect concentration.

This causes a low bias of about 6% to 15%. The agency states this could lead to incorrect patient classification, delays in treatment, and errors in clinical treatment decisions, which may cause patient harm.

The reagent and standard set are used in laboratory tests. The products were distributed nationwide in the United States, as well as worldwide, including Puerto Rico. The company began the recall on October 21, 2026.

Is your set affected?

2 products

Match the UDI-DI and lot code listed under your product.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Infinity Ammonia Standard Set

    Product details from the notice

    Infinity Ammonia Standard Set Model/Catalog Number: TR60901 Product Description: In Vitro Diagnostic calibrator and control

    UDI-DI
    • 00845275003314
    Lot code
    • 382963
    • 364987
    • 307166
  2. Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers

    Product details from the notice

    Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers Model/Catalog Number: OSR61154 Product Description: In Vitro Diagnostic reagent, calibrator, and control

    UDI-DI
    • 00845275000016
    Lot code
    • 2915
    • 2916
    • 9511
    • 9510
    • 0644
    • 0645
    • 0614
    • 0613
    • 5301
    • 9901
    • 2525
    • 7602
    • 7603
    • 4111
    • 2393
    • 7411
    • 6624
    • 3223
    • 3231
    • 7779
    • 7780
    • 0206
    • 0209

Recall details

Posted by FDA
Recall began
Status
Open
Sold in

Worldwide - US Nationwide distribution in the states and territory of California, Illinois, Pennsylvania, Puerto Rico and the countries of Australia, GER, South Korea.

Recalled by
Fisher DiagnosticsMiddletown, Virginia
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

FDA event number
100164
Recall numbers
Z-0533-2027 to Z-0534-2027
Recall type
Started by the company
How the company first told customers
More than one way: email, fax, letter, press release, phone or visit

All recalls from Fisher Diagnostics

Timeline

FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.

  1. Fisher Diagnostics began the recall

  2. FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)

  3. Added to Recall Canary

Summary by Recall Canary, written with AI · summary updated Nov 25, 2026 · Official source: U.S. Food and Drug Administration (opens in a new tab) recall · cite: “Fisher Diagnostics Infinity Ammonia Reagent and Standard Set Recall.” Recall Canary. https://recallcanary.com/recall/fda/fisher-diagnostics-infinity-ammonia-standard-set.