FDA recall, , Moderate risk
General Medical Merate Discovery RF180GMM
Discovery RF180GMM systems with Thales RF or FL4 detectors do not display the air kerma rate in a specific setting.
What to do
Check yoursIf a code on your device is listed below:
- Check the lot against the recall notice and ask your provider or supplier before you stop using it.
Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.
What happened
General Medical Merate S.p.A., of Seriate, Italy, is recalling Discovery RF180GMM systems equipped with Thales RF or FL4 detectors. The recall is voluntary and was requested by the FDA. The system software does not display the air kerma rate (AKR) when the device is operated in a specific setting. That setting uses the detector's Logical Mode 11 at an image acquisition frequency greater than six images per second.
The company began the recall on September 16, 2026. The products were distributed nationwide. The FDA added the recall to its Enforcement Report and classified it on November 25, 2026.
Is yours affected?
Affected if its serial number is listed.
Discovery RF180GMM equipped with Thales RF or FL4 detectors
- Serial number
- 01129
- 01946
Recall details
- Posted by FDA
- Recall began
- Status
- Open
- Weekly archive
- Listed the week of Nov 23, 2026
- Sold in
Distributed nationwide.
- Recalled by
- General Medical Merate S.p.A.Seriate, Italy
- Hazard
- Software defect
- Risk
Moderate risk
Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.
FDA Class II recall · How risk is rated
- FDA event number
- 100227
- Recall number
- Z-0551-2027
- Recall type
- At FDA's request
All recalls from General Medical Merate S.p.A.
Timeline
FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.
General Medical Merate S.p.A. began the recall
FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)