PALAMIX uno Mixing Unit

Heraeus recalls PALAMIX uno mixing units: mixing rod may loosen in units over 17 months old; shelf life reduced to 1.5 years.

What to do

Check the age of your PALAMIX uno unit. If it is older than 17 months, stop using it and contact Heraeus Medical GmbH for a replacement or further instructions.

Recalled
Listed
Updated
Recall No.
Z-1997-2026
Agency
FDA
Firm
Heraeus Medical GmbH (Dental Division)
Recall ID
98608
Status
ONGOING

Why it was recalled

Heraeus Medical GmbH is recalling the PALAMIX uno mixing unit (Material Number 66057893) because long-term stability testing determined that in units older than 1.5 years (17 months), the mixing rod may, in rare cases, come loose from its lock during the mixing process. The shelf life is being reduced from 3 years to 1.5 years (17 months).

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

ProductRecall #UDI-DI
Heraeus, PALAMIX uno. Material Number: 66057893.Z-1997-202666057893, 04260102133587
Heraeus, PALAMIX duo. Material Number: 66057897.Z-1998-202666057897, 04260102133617

Where it was sold

Distributed nationwide.