Skip to main content

FDA recall, , Moderate risk

ICU Medical Plum Duo and Plum Solo Infusion Pumps

ICU Medical Plum Duo and Plum Solo Infusion Pumps can prevent programming a piggyback flush, causing a violation message.

What happened

ICU Medical, Inc. of Lake Forest, Illinois recalled the Plum Solo Precision IV Pump, the Plum Duo Precision IV Pump, and the Plum Duo Infusion Pump beginning December 15, 2025. The recall was added to the FDA's Enforcement Report and classified on January 16, 2026. The pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy.

The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. Distribution was worldwide, including the US Nationwide and the country of Philippines.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

3 products

Affected if its UDI-DI is listed.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)

    UDI-DI
    • 400020401
    • 400021001
    Affected units
    400020401 (US), Software Versions: 1.1.1, 1.1.2, 1.1.3, 1.1.4 400021001 (Philippines), Software Versions: 1.1.0
  2. ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

    UDI-DI
    • 400010401
    • 400011001
    Affected units
    400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0
  3. ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)

    UDI-DI
    • 400020403
    • 400021003
    Affected units
    400020403 (US), Software Versions: 1.2.2; 400021003 (Philippines), Software Versions: 1.2.0

Recall details

Company began the recall
Sold in

Worldwide distribution - US Nationwide and the country of Philippines.

Company
ICU Medical, Inc.Lake Forest, Illinois
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

Event ID
98128
Recall numbers
Z-1118-2026 to Z-1120-2026

All recalls from ICU Medical, Inc.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. In FDA's weekly enforcement report

  4. Listed on Recall Canary

Summary by Recall Canary, written with AI · summary updated Oct 23, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab)