Skip to main content

FDA recall, , Moderate risk

ICU Medical Pulsator Arterial Blood Sampling Kit and Portex Spinal Tray

Pulsator Arterial Blood Sampling Kit and Portex Spinal Tray may have cracked syringe collars, leading to delayed or repeated blood draw or fluid leakage.

What happened

ICU Medical Inc. of Minneapolis, Minnesota, recalled Pulsator Arterial Blood Sampling Kit and Portex Spinal Tray because affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw or fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.

The recall began December 19, 2025, and was added to the FDA's Enforcement Report on January 15, 2026. Distribution was in California, Florida, Illinois, Indiana, Massachusetts, Michigan, Minnesota, Missouri, Nebraska, New Jersey, New York, Ohio, Oklahoma, Texas and Washington.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is your kit affected?

2 products

Affected if its UDI-DI and lot code are listed under that product.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG

    UDI-DI
    • 30351688073062
    • 30351688006176
    • 30351688029540
    • 30351688039150
    • 30351688081173
    Lot code
    • 6141992
    • 6156228
    • 4080P-2
    • 6141999
    • 6142000
    • 6147906
    • 4084P-1
    • 6147907
    • 4080PE
    • 6131839
    • 4080PG
    • 6131838
    • 6142001
    • 6142002
  2. Portex Spinal Tray, Item No. A3729-24 A3595

    UDI-DI
    • 30351688075691
    • 35019517108211
    Lot code
    • 6146204
    • A3595
    • 6157980

Recall details

Company began the recall
Company
ICU Medical Inc.Minneapolis, Minnesota
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

Event ID
98168
Recall numbers
Z-1100-2026 to Z-1101-2026

All recalls from ICU Medical Inc.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. In FDA's weekly enforcement report

  4. Listed on Recall Canary

Summary by Recall Canary, written with AI · summary updated Oct 23, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab)