Skip to main content
Sunday email

FDA recall, , Moderate risk

Inpeco FlexLab X Automation Track

FlexLab X power cables may have loose electrical connections, which could injure laboratory staff.

Product photo

What to do

Check yours

If a code on your device is listed below:

  1. Check the lot against the recall notice and ask your provider or supplier before you stop using it.

Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.

What happened

Inpeco S.A., of Lugano, Switzerland, is recalling 35 units of its FlexLab X system, model FXX. Certain power cables used in the system may have loose electrical connections. This could affect one or more of the cable's internal wires. The system may malfunction, causing delays in sample processing, or potentially injuring laboratory staff.

The affected parts are the power cable end plug and end socket at the automation track head and at the T turn and L turn modules, and the central power junction box at the automation track and track T module. The loose connections are due to screws or terminals that were not tightened to the required level.

The units were distributed nationwide in the US and in other countries, including Denmark, France, Japan, and others. The company started the recall on its own and told customers by letter, beginning October 19, 2026.

Is your box affected?

Compare the details below with your product and check the FDA notice to confirm whether it is affected.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Product details

    FlexLab X;
    Model/Version: FXX;
    Component (Description): 1. MEE0074 (Power Cable End Plug at the Automation Track Head, at T Turn Module and L Turn Module), 2. MEE0075 (Power Cable End Socket at the Automation Track Head, at T Turn Module and L Turn Module), 3. MEE0082 (Power Cable Central Power Junction Box at the Automation Track and Track T Module);

    UDI-DI
    • 07640172344002
    Serial number
    • FXX.0062
    • FXX.0065
    • FXX.0067
    • FXX.0068
    • FXX.0069
    • FXX.0070
    • FXX.0071
    • FXX.0072
    • FXX.0073
    • FXX.0074
    • FXX.0075
    • FXX.0076
    • FXX.0077
    • FXX.0082
    • FXX.0083
    • FXX.0084
    • FXX.0085
    • FXX.0086
    • FXX.0087
    • FXX.0088
    • FXX.0089
    • FXX.0090
    • FXX.0092
    • FXX.0093
    • FXX.0098
    • FXX.0099
    • FXX.0100
    • FXX.0101
    • FXX.0103
    • FXX.0104
    • FXX.0105
    • FXX.0106
    • FXX.0107
    • FXX.0108
    • FXX.0110

Recall details

Posted by FDA
Recall began
Status
Open
Sold in

Distributed in Missouri and in Denmark, Egypt, France, Greece, Hungary, Italy, Japan, Malaysia, Netherlands, Romania, Saudi Arabia, Spain, Switzerland and Turkey.

Recalled by
Inpeco S.A.Lugano, Switzerland
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

FDA event number
100107
Recall number
Z-0550-2027
Quantity recalled
35 units;
Recall type
Started by the company
How the company first told customers
Letter

All recalls from Inpeco S.A.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.

  1. Inpeco S.A. began the recall

  2. FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)

  3. Added to Recall Canary

Summary by Recall Canary, written with AI · summary updated Nov 26, 2026 · Official source: U.S. Food and Drug Administration (opens in a new tab) recall · cite: “Inpeco FlexLab X Automation Track Recall.” Recall Canary. https://recallcanary.com/recall/fda/inpeco-sa-flexlab-x.