Skip to main content

Ongoing FDAEvent 98944 (3 months ago)

Medacta Short Monobloc Stem Broaches

Short Monobloc Stem Broaches may increase the risk of fracture in the medial region of the proximal humerus.

Moderate risk

Using it may cause temporary or medically reversible health problems. Serious harm is unlikely.

FDA Class II recall · How risk is rated

What to do

What happened

Medacta USA Inc., of Memphis, Tennessee, began a recall on April 30, 2026, for Short Monobloc Stem Broaches, part of the Medacta Shoulder System.

The surgical instruments, used during shoulder arthroplasty procedures, do not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.

The FDA added the recall to its Enforcement Report on June 29, 2026. Distribution was nationwide in the states of CO, AR, IN, PA, UT, and FL.

Summary by Recall Canary from the FDA notice.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Affected if its UDI-DI and lot code are listed under that product.

  1. Short Monobloc Stem Broaches, Size/REF

    UDI-DI
    • 07630542768910
    • 07630542768927
    • 07630542768934
    • 07630542768941
    • 07630542768958
    • 07630542768965
    • 07630542768972
    • 07630542768989
    • 07630542768996
    • 07630542769009
    Lot
    • 04.01.10.0970
    • 2585086
    • 04.01.10.0971
    • 2585087
    • 04.01.10.0972
    • 2585088
    • 04.01.10.0973
    • 2585089
    • 04.01.10.0974
    • 2585090
    • 04.01.10.0975
    • 2585091
    • 04.01.10.0976
    • 2585092
    • 04.01.10.0977
    • 2585093
    • 04.01.10.0978
    • 2585094
    • 04.01.10.0979
    • 2585095

Details

Company began the recall
Sold in

US Nationwide distribution in the states of Colorado, Arkansas, Indiana, Pennsylvania, Utah, Florida.

Event ID
98944
Recall No.
Z-2618-2026
Company
Medacta Usa IncMemphis, Tennessee
  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. Listed on Recall Canary

  4. In FDA's weekly enforcement report

  5. Summary updated

Summary by Recall Canary, written with AI · summary updated Oct 3, 2026 · primary source: U.S. Food and Drug Administration