Velys Enabling Tech Robotic-Assist Station
Medos recalls Velys Robotic-Assist Station over software issue causing unintended arm movement.
If you have the Velys Enabling Tech Robotic-Assist Station, stop using it and contact Medos International Sarl for instructions on updating the software or returning the device.
Recall at a glance
- Quantity recalled
- 29 stations
Why it was recalled
Medos International Sarl is recalling the Velys Enabling Tech Robotic-Assist Station (REF: 451611001) due to a software issue in the robotic controller that may cause unintended movement of the robotic arm during bone preparation or instrument positioning.
Is mine affected?
Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.
Velys Enabling Tech Robotic-Assist Station REF: 451611001
- UDI-DI
- 10705034645995