Skip to main content
Recall Canary
munevo DRIVE Powered Wheelchair Control System
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

munevo DRIVE Powered Wheelchair Control System

Munevo recalls wheelchair control systems over software calibration issue causing steering drift and collision risk.

What to do

Stop using the affected munevo DRIVE or munevo DRIVE r-net system immediately. Contact Munevo GmbH for a software update to correct the calibration issue.

Recalled
Listed
Updated
Recall No.
Z-2856-2026
Agency
FDA
Firm
MUNEVO GMBH
Recall ID
99328
Status
ONGOING

Why it was recalled

Munevo GmbH is recalling the munevo DRIVE and munevo DRIVE r-net powered wheelchair hands-free control application and smart glasses interface due to a software issue that may cause the wheelchair's driving mechanism to fall out of calibration. This can result in unexpected steering drift or imprecise control, leading to veering from the intended path and potential low-speed collision, tipping, or injury.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System

UDI-DI
04260691380010

Where it was sold

Distributed nationwide.

Timeline

  1. Firm initiated recall
  2. Publicly announced
  3. Listed on Recall Canary
  4. Summary updated

Other Software Defect recalls

← Back to all recalls

Editorial summary by Recall Canary · summary updated 2026-08-02 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification