Skip to main content

FDA recall, , Moderate risk

Natus Embletta MPR PG Sleep Monitor

The Embletta MPR PG Multi Parameter Recorder Sleep Monitor may be unable to correctly format its internal memory.

What happened

Natus Medical Incorporated of Oakville, Canada, voluntarily initiated a recall of the Embletta MPR PG Multi Parameter Recorder Sleep Monitor on September 29, 2026. The device may be unable to correctly format its internal memory, which can cause previously recorded studies to remain on the device and be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely.

The recall was added to the FDA's Enforcement Report and classified on October 30, 2026. The device was distributed worldwide, including the US nationwide and the countries of Canada, China, Italy, and Portugal. Customers were notified by letter.

The company initiated the recall voluntarily for 288 units.

What to do

The FDA notice gives no instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Match the model number and UDI-DI listed under your product.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Embletta MPR PG Multi Parameter Recorder Sleep Monitor

    Model number
    • 2003011
    UDI-DI
    • 00382830023764

Recall details

Posted by FDA
Recall began
Status
Open
Sold in

US Nationwide and the countries of Canada, China, Italy, Portugal.

Recalled by
NATUS MEDICAL INCORPORATEDOakville, Canada
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

FDA event number
99941
Recall number
Z-0319-2027
Quantity recalled
288 units
Recall type
Voluntary: Firm initiated
How the company first told customers
Letter

All recalls from NATUS MEDICAL INCORPORATED

Timeline

FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.

  1. NATUS MEDICAL began the recall

  2. FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)

  3. Added to Recall Canary

Summary by Recall Canary, written with AI · summary updated Oct 31, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab)