Skip to main content

Ongoing FDAEvent 99704 (13 days ago)

Steris VERIFY Biological Steam Test Packs and Biological Indicators

Verify Biological/Integrator Challenge Pack may have spore population below specification, potentially showing no growth in steam sterilization monitoring.

Moderate risk

Using it may cause temporary or medically reversible health problems. Serious harm is unlikely.

FDA Class II recall · How risk is rated

What to do

What happened

Steris Corporation recalled certain Verify Biological/Integrator Challenge Pack and Verify Dual Species Self-Contained Biological Indicator and Verify Biological Steam Test Pack products because the Geobacillus stearothermophilus spore population may fall below minimum shelf life specification.

When used to monitor steam sterilization cycles, the biological indicator may show no growth, indicating the steam cycle was effective even if it did not achieve the intended spore reduction. The recall began August 31, 2026.

Summary by Recall Canary from the FDA notice.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Match your product and all the details in one listed combination.

  1. VERIFY Biological Steam Test Pack Model/

    Lot
    • 260810
    1. UDI-DI
      • 10724995220300
      Catalog code
      • S3065
    2. UDI-DI
      • 10724995023864
      Catalog code
      • S3069
  2. VERIFY Biological/Integrator Challenge Pack Model/

    Lot
    • 260810
    1. UDI-DI
      • 10724995220324
      Catalog code
      • LCB006
    2. UDI-DI
      • 10724995220331
      Catalog code
      • LCB007
  3. VERIFY Dual Species Self-Contained Biological Indicator Model/

    Lot
    • 260810
    1. UDI-DI
      • 50724995023824
      Catalog code
      • S3061
    2. UDI-DI
      • 10724995023833
      Catalog code
      • S3060

Details

Company began the recall
Event ID
99704
Company
Steris Corporation
  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. Listed on Recall Canary

  4. In FDA's weekly enforcement report

All recalls from Steris Corporation

Summary by Recall Canary, written with AI · summary updated Oct 3, 2026 · primary source: U.S. Food and Drug Administration