Tandem Diabetes Care Tandem Mobi Insulin Pump

Tandem Mobi insulin pumps may generate false-positive occlusion alarms, suspending insulin delivery and potentially causing elevated blood glucose levels.

What happened

Tandem Diabetes Care, Inc., of San Diego, California, is voluntarily recalling certain Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, 1018078. The recall began September 23, 2026, and covers 80,938 pumps (77,559 US, 3,379 OUS). The FDA added it to the Enforcement Report and classified it on October 30, 2026.

Tandem has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. Distribution is nationwide in the US and worldwide including Canada.

What to do

The FDA notice gives no instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Affected if its model number and UDI-DI are listed under that product.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Tandem Mobi Insulin Pump

    Model number
    • 1004004
    • 1010750
    • 1012676
    • 1012719
    • 1013655
    • 1013656
    • 1013501
    • 1013700
    • 1014081
    • 1014801
    • 1014811
    • 1014813
    • 1014936
    • 1016911
    • 1016912
    • 1016913
    • 1016914
    • 1018075
    • 1018077
    • 1018078
    UDI-DI
    • 00389152075013
    • 00389152271910
    • 00389152350110
    • 00389152365510
    • 00389152365619
    • 00389152370019
    • 00389152480114
    • 00389152481111
    • 00389152481319

Recall details

Posted by FDA
Recall began
Status
Open
Sold in

Worldwide - US Nationwide distribution including in the states of CA, FL, NC, OK, UT, WA, IL, PA, MN, VA, NY, IN, HI, OR, GA, MI, KS, CO, WI, SC, null, TX, NE, AZ, OH, NM, NV, MA, SD, IA, TN, ID, MD, RI, CT, NJ, ME, KY, MT, WV, VT, MS, AR, AL, DC, LA, MO, NH, PR, ND, AK, DE, WY and the country of Canada.

Recalled by
Tandem Diabetes Care, Inc.San Diego, California
Risk

High risk

There is a reasonable chance that using it could cause serious health problems or death.

FDA Class I recall · How risk is rated

FDA event number
99851
Recall number
Z-0316-2027
Quantity recalled
80,938 (77,559 US, 3,379 OUS)
Recall type
Voluntary: Firm initiated

All recalls from Tandem Diabetes Care, Inc.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.

  1. Tandem Diabetes Care began the recall

  2. FDA posted the recall to its Enforcement Report and rated it Class I (high risk)

  3. Added to Recall Canary

Summary by Recall Canary, written with AI · summary updated Oct 31, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab)