FDA recall, , Moderate risk
Vortex Surgical Ophthalmic Surgical Instruments
Volk Single Use Vitrectomy Lenses may have voids in the seal of Tyvek pouches, which could lead to infection.
What happened
Vortex Surgical is recalling 13 products.
What to do
We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:
- A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
- Talk to your health care provider before you stop using it.
Is your kit affected?
Compare the details below with your product and check the FDA notice to confirm whether it is affected.
A UDI-DI is the device identifier printed with the barcode on the device label.
Vortex Surgical I2 Injection Kit, VS0500
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- UDI-DI
- 810123480920
- 810123481330
- Lot code
- 2411012
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.25GA Convenience Kit - XES VS0644.25-XES
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- UDI-DI
- 810123482917
- 810123482924
- 810123482429
- 810123482443
- 810123482412
- 810123482436
- Lot code
- 2410055
- 25GA
- 2411028
- 2410057
Rumex Disposable Diamond Dusted ILM Elevator, 12-7523
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- UDI-DI
- 5060720920298
- Lot code
- 2411033
Vortex Surgical TID Pharos Illuminated Depressor, VS0801B
Reason: XXX
- UDI-DI
- 810123481309
- 810123481712
- Lot code
- 2403027R
Product details
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated Flex-Tip VK401213 3. 23G Laser Probe OS4 Intuitive Exendable VK401323 4. 25G Laser Probe OS4 Flex-Tip VK401115 5. 25G Laser Probe OS4 Illuminated Flex-Tip VK401215 6. 25G Laser Probe OS4 Intuitive Exendable VK401325 7. 27G Laser Probe OS4 Intuitive Exendable VK401327
- UDI-DI
- 810123483716
- 810123483792
- 810123483730
- 810123483815
- 810123483761
- 810123483846
- 810123483723
- 810123483808
- 810123483747
- 810123483822
- 810123483778
- 810123483853
- 810123483785
- 810123483860
- Lot code
- 2410031
- 23G
- OS4
- 2410034
- 2410033
- 25G
- 2410032
- 2410058
- 2411022
- 2410035
- 2412025
- 27G
- 2411021
25GA Subretinal Injection Cannula VS0220.25
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- UDI-DI
- 810123480418
- 810123480692
- Lot code
- 2410043
Vortex Surgical Lindsell Sutured IOL Marker, VS0390 - Scleral Marker
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- UDI-DI
- 810123483655
- 810123483662
- Lot code
- 2408035
Product details
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Vortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3. 25GA Laser Probe Flex-Tip Cat No. VS0130.25 4. 25GA Laser Probe Illuminated Curved Cat No. VS0125.25B 5. 25GA Laser Probe MaxReach Cat No. VS0140.25 6. 25GA Laser Probe Straight Cat No. VS0110.25
- UDI-DI
- 810123480036
- 810123480180
- 810123480043
- 810123480197
- 810123480111
- 810123480265
- 810123480098
- 810123480241
- 810123480135
- 810123480289
- 810123480012
- 810123480166
- Lot code
- 2411027
- 25GA
- 2411026
- 2411024
- 2411031
- 2411025
- 2410029
- 2411023
- 2411030
Product details
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Vortex Disposable Forceps and Cannula 1. 23GA ACTU8 Forceps Adaptive VS0740.23 2. 25GA ACTU8 Forceps Adaptive VS0740.25 3. 25GA ACTU8 Forceps ILM VS0744.25 4. 25GA ACTU8 Forceps ILM -Myopic VS0744.25-35 5. 25GA ACTU8 Forceps Pro-Grip VS0713.25 6. 25GA ACTU8 Talon Forceps ILM VS0752.25 7. 25GA ACTU8 Talon Forceps Pro-Grip VS0751.25 8. 27GA ACTU8 Forceps ILM VS0744.27 9. 23GA Soft Tip Cannula VS0200.23 10. 23GA Soft Tip Cannula - 0.5mm VS0205.23 11. 25GA Soft Tip Cannula VS0200.25 12. 25GA Soft Tip Cannula - 0.5mm VS0205.25 13. 25GA Soft Tip Cannula - 0.5mm - Bulk VS0205.25B 14. 25GA Soft Tip Cannula-Bulk VS0200.25B 15. 27GA Soft Tip Cannula VS0200.27
- UDI-DI
- 810123481125
- 810123481538
- 810123481132
- 810123481545
- 810123481231
- 810123481644
- 810123481248
- 810123481651
- 810123481101
- 810123481514
- 810123481835
- 810123481859
- 810123482375
- 810123482405
- 810123481255
- 810123481668
- 810123480302
- 810123480562
- 810123480388
- 810123480661
- 810123480319
- 810123480586
- 810123480395
- 810123480678
- 810123481927
- 810123480593
- 810123480326
- 810123480609
- Lot code
- 2411009
- 25GA
- ACTU8
- 2410051
- 2410021
- 2410053
- 2411006
- 2410059
- 2410050
- 2411011
- 2411003
- 2410060
- 27GA
- 2410054
- 23GA
- 2410037
- 0.5mm
- 2411032
- 2410039
- 2410041
- 2410042
- 2411018
- 2410040
- 2412026
Tecfen Retractable Membrane Polisher, QTPR1267-23
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- UDI-DI
- 817618024508
- Lot code
- 2411013
Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy Lens, Catalog VFD 2. Volk Magnifying Lens, Catalog VMD
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- UDI-DI
- 810123482290
- 810123482306
- 810123482825
- 810123482832
- Lot code
- 2410036
- 2411020
- 2411019
Vortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- UDI-DI
- 810123480470
- 810123480753
- 810123480517
- 810123480791
- Lot code
- 2411017
- 2411016
Vortex Surgical 25GA I.D.D. Internal Delivery Device, VS0250.25
Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots.
Rest of the reason
A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- UDI-DI
- 810123480456
- 810123480739
- Lot code
- 2410044
Recall details
- Company began the recall
- Sold in
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
- Company
- Vortex Surgical Inc.Saint Charles, Missouri
- Hazard
- Sterility problem
- Risk
Moderate risk
Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.
FDA Class II recall · How risk is rated
- Event ID
- 98255
- Recall numbers
- Z-1244-2026 to Z-1256-2026
Timeline
FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.
Company began the recall
Added to FDA's Enforcement Report · Classified by FDA
In FDA's weekly enforcement report
Listed on Recall Canary