Skip to main content

Ongoing FDAEvent 99834 (20 days ago)

Philips MRI Scanners and Upgrade Kits

Ingenia Blue Seal SE and Blue Seal XE export incorrect diffusion data when using Enhanced MR Dicom format.

Moderate risk

Using it may cause temporary or medically reversible health problems. Serious harm is unlikely.

FDA Class II recall · How risk is rated

What to do

What happened

Philips North America recalled multiple MR systems, including the Ingenia Blue Seal SE and Blue Seal XE, because when exporting diffusion tensor imaging (DTI) scans using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the export is used with third-party software for tractography, such as mapping nerve fiber pathways in the brain, the software will not correctly process the DTI data for reconstruction. The recall began August 28, 2026.

Summary by Recall Canary from the FDA notice.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Compare the details below with your product and check the FDA notice to confirm whether it is affected.

  1. Upgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838104402
  2. MR 5300; Model number (REF): 1. 782110, 2. 782152, 3. 782156, 4. 782196; Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838099364
    • 00884838112841
    • 00884838115675
    • 00884838128446
  3. Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838055339
    • 00884838009813
    • 00884838098299
    • 00884838068452
  4. Upgrade dStream; Model number (REF): 782127; Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838105591
  5. Product details

    Ingenia Elition S and Ingenia Elition X.;
    Model number (REF): 1. 781357 (S), 2. 782106 (S), 3. 782137 (S), 4. 782157 (S), 5. 781358 (X), 6. 782107 (X), 7. 782119 (X), 8. 782136 (X), 9. 782158 (X), 10. 782198 (X);
    Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838088108
    • 00884838098329
    • 00884838108615
    • 00884838115682
    • 00884838088115
    • 00884838098336
  6. Ingenia Blue Seal SE and Blue Seal XE; Model number (REF): 1. 782185 (SE), 2. 782192 (XE); Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838125698
    • 00884838128422
  7. Product details

    SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, and SmartPath to Ingenia Elition X;
    Model number (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146 (1.5T), 4. 782165 (1.5T), 5. 782145 (3.0T), 6. 782164 (3.0T), 7. 781270 (XR and 3.0T), 8. 782113 (XR and 3.0T), 9. 782129 (XR and 3.0T), 10. 782118 (Ingenia Elition X), 11. 782144 (Ingenia Elition X), 12. 782163 (Ingenia Elition X);
    Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838095076
    • 00884838098886
    • 00884838108691
    • 00884838115767
    • 00884838108684
    • 00884838115750
    • 00884838095083
    • 00884838098909
    • 00884838105805
  8. SmartPath to MR 7700; Model number (REF): 1. 782178, 2. 782207; Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838122239
    • 00884838128576
  9. MR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838104112
    • 00884838112858
    • 00884838115668
    • 00884838128439
  10. Product details

    Ingenia 1.5T, 1.5T CX, and 1.5T S;
    Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262 (1.5T CX), 7. 781347 (1.5T S);
    Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838099715
    • 00884838055322
    • 00884838009820
    • 00884838099043
    • 00884838108646
    • 00884838068445
  11. Product details

    Ingenia Ambition S and Ingenia Ambition X.;
    Model number (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 782159 (S), 5. 782194 (S), 6. 781356 (X), 7. 782109 (X), 8. 782138 (X), 9. 782160 (X), 10. 782195 (X);
    Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838090057
    • 00884838098343
    • 00884838108639
    • 00884838115705
    • 00884838128477
    • 00884838090040
    • 00884838098350
  12. Product details

    Evolution upgrade 1.5T amd Evolution upgrade 3.0T;
    Model number (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782166 (1.5T), 4. 782201(1.5T), 5. 782117 (3.0T), 6. 782143 (3.0T), 7. 782162 (3.0T);
    Software Versions: R11, R12 and R13;

    UDI-DI
    • 00884838099722
    • 00884838108714
    • 00884838115774
    • 00884838128552
    • 00884838099739
    • 00884838108660
    • 00884838115736

Details

Company began the recall
Sold in

Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.

Event ID
99834
Recall numbers
Z-3307-2026 to Z-3318-2026
Company
Philips North AmericaCambridge, Massachusetts
  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. Listed on Recall Canary

  4. Summary updated

  5. In FDA's weekly enforcement report

All recalls from Philips North America

Summary by Recall Canary, written with AI · summary updated Oct 3, 2026 · primary source: U.S. Food and Drug Administration