Philips Spectral CT 7500
The Spectral CT 7500 software may crash during bolus tracking.
Moderate risk
Using it may cause temporary or medically reversible health problems. Serious harm is unlikely.
FDA Class II recall · How risk is rated
What happened
Philips North America Llc is recalling the Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333. The Bolus tracking scan application may crash with a system error message, and Bolus Tracker shots may continue even if the contrast threshold is reached, potentially causing additional radiation to the patient. The recall began on August 18, 2026, and was added to the FDA's Enforcement Report on September 29, 2026.
Summary by Recall Canary from the FDA notice.
What to do
We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:
- A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
- Talk to your health care provider before you stop using it.
Is yours affected?
Affected if its UDI-DI matches.
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
- UDI-DI
- 00884838101111
Details
- Company began the recall
- Sold in
U.S. and U.S. Territories.
- Event ID
- 99865
- Recall No.
- Z-3155-2026
- Company
- Philips North America LlcCambridge, Massachusetts
- Hazard
- Device defect
Company began the recall
Added to FDA's Enforcement Report · Classified by FDA
Listed on Recall Canary
Summary updated
In FDA's weekly enforcement report