Skip to main content

FDA recall, , Moderate risk

Roche Cobas C 303 Clinical Chemistry Analyzer

Roche cobas c 303 Clinical Chemistry Analyzer has vacuum tubing in its cell rinse unit that deteriorates prematurely.

What to do

Check yours

If a code on your device is listed below:

  1. Check the lot against the recall notice and ask your provider or supplier before you stop using it.

Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.

What happened

Roche Diagnostics Operations, Inc. is recalling 1002 units of its Roche cobas c 303 analytical unit, a clinical chemistry analyzer. Premature deterioration of the vacuum tubing in the cell rinse unit occurs due to excess bending stress. The tubing may cause pinholes, cracks or tubing rupture, and fluid may drop into random reaction cells without any system alarm.

This is increasing the risk of erroneous patient results. The company started the recall on its own. Customers were told by letter.

The recall began on September 23, 2026.

Distribution was nationwide and included a territory. The FDA added the recall to its Enforcement Report on November 18, 2026.

Is yours affected?

Affected if its UDI-DI matches.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Roche cobas c 303 analytical unit

    Product details from the notice

    Roche cobas c 303 analytical unit, Material Number 09793194001, clinical chemistry analyzer

    UDI-DI
    • 07613336220153
    Affected units
    All units up to and including SN (Serial Number) 7696-10

Recall details

Posted by FDA
Recall began
Status
Open
Sold in

Distributed nationwide and in Puerto Rico.

Recalled by
Roche Diagnostics Operations, Inc.Indianapolis, Indiana
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

FDA event number
99980
Recall number
Z-0481-2027
Quantity recalled
1002 units
Recall type
Started by the company
How the company first told customers
Letter

All recalls from Roche Diagnostics Operations, Inc.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.

  1. Roche Diagnostics Operations, Inc. began the recall

  2. FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)

  3. Added to Recall Canary

Summary by Recall Canary, written with AI · summary updated Nov 20, 2026 · Official source: U.S. Food and Drug Administration (opens in a new tab) recall · cite: “Roche Cobas C 303 Clinical Chemistry Analyzer Recall.” Recall Canary. https://recallcanary.com/recall/fda/roche-cobas-c-303-analytical-unit.