FDA recall, , Moderate risk
Roche Cobas C 303 Clinical Chemistry Analyzer
Roche cobas c 303 Clinical Chemistry Analyzer has vacuum tubing in its cell rinse unit that deteriorates prematurely.
What to do
Check yoursIf a code on your device is listed below:
- Check the lot against the recall notice and ask your provider or supplier before you stop using it.
Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.
What happened
Roche Diagnostics Operations, Inc. is recalling 1002 units of its Roche cobas c 303 analytical unit, a clinical chemistry analyzer. Premature deterioration of the vacuum tubing in the cell rinse unit occurs due to excess bending stress. The tubing may cause pinholes, cracks or tubing rupture, and fluid may drop into random reaction cells without any system alarm.
This is increasing the risk of erroneous patient results. The company started the recall on its own. Customers were told by letter.
The recall began on September 23, 2026.
Distribution was nationwide and included a territory. The FDA added the recall to its Enforcement Report on November 18, 2026.
Is yours affected?
Affected if its UDI-DI matches.
A UDI-DI is the device identifier printed with the barcode on the device label.
Roche cobas c 303 analytical unit
Product details from the notice
Roche cobas c 303 analytical unit, Material Number 09793194001, clinical chemistry analyzer
- UDI-DI
- 07613336220153
- Affected units
- All units up to and including SN (Serial Number) 7696-10
Recall details
- Posted by FDA
- Recall began
- Status
- Open
- Weekly archive
- Listed the week of Nov 16, 2026
- Sold in
Distributed nationwide and in Puerto Rico.
- Recalled by
- Roche Diagnostics Operations, Inc.Indianapolis, Indiana
- Hazard
- Device defect
- Risk
Moderate risk
Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.
FDA Class II recall · How risk is rated
- FDA event number
- 99980
- Recall number
- Z-0481-2027
- Quantity recalled
- 1002 units
- Recall type
- Started by the company
- How the company first told customers
- Letter
All recalls from Roche Diagnostics Operations, Inc.
Timeline
FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.
Roche Diagnostics Operations, Inc. began the recall
FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)
Added to Recall Canary