Ongoing FDAEvent 99731 (20 days ago)

United Imaging Computed Tomography and Positron Emission Tomography Systems

The systems have a software problem when Split or Local function is enabled and multiple RIS exams are merged into a single study.

Moderate risk

Using it may cause temporary or medically reversible health problems. Serious harm is unlikely.

FDA Class II recall · How risk is rated

What to do

What happened

Shanghai United Imaging Healthcare Co., Ltd. recalled Computed Tomography X-ray Systems and Positron Emission Tomography and Computed Tomography Systems because a software problem, when Split or Local function is enabled and multiple RIS exams are merged into a single study, prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.

The recall began August 11, 2026, and was added to the FDA's Enforcement Report on September 25, 2026.

Summary by Recall Canary from the FDA notice.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Affected if its UDI-DI is listed.

  1. Positron Emission Tomography and Computed Tomography System

    Model
    uMI
  2. Computed Tomography X-ray System

    Model
    uCT
  3. Computed Tomography X-ray System, Model: uCT 530 with software

    UDI-DI
    • 06971576831012
  4. Computed Tomography X-ray System, Model: uCT 760 with software

    UDI-DI
    • 06971576831043
  5. Positron Emission Tomography and Computed Tomography System, Model: uMI 780

    UDI-DI
    • 06971576832033
  6. Computed Tomography X-ray System, Model: uCT 550 with software

    UDI-DI
    • 06971576831036
  7. Positron Emission Tomography and Computed Tomography System, Model: uMI Panorama 35 with software

    UDI-DI
    • 06971576832170
  8. Computed Tomography X-ray System, Model: uCT ATLAS Astound with software

    UDI-DI
    • 06971576831340
  9. Positron Emission Tomography and Computed Tomography System, Model: uMI Panorama GS with software

    UDI-DI
    • 06971576832187
  10. Positron Emission Tomography and Computed Tomography System, Model: uMI Panvivo S with software

    UDI-DI
    • 06971576832255
  11. Computed Tomography X-ray System, Model: uCT 780 with software

    UDI-DI
    • 06971576831050
  12. Positron Emission Tomography and Computed Tomography System, Model: uMI Panvivo with software

    UDI-DI
    • 06971576832231
  13. Computed Tomography X-ray System, Model: uCT ATLAS with software

    UDI-DI
    • 06971576831104
  14. Positron Emission Tomography and Computed Tomography System, Model: uMI 550 with software

    UDI-DI
    • 06971576832026

Details

Company began the recall
Event ID
99731
Recall numbers
Z-3277-2026 to Z-3288-2026
Company
Shanghai United Imaging Healthcare Co., Ltd.Shanghai, China
  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. Listed on Recall Canary

  4. Summary updated

  5. In FDA's weekly enforcement report

Summary by Recall Canary, written with AI · summary updated Oct 3, 2026 · primary source: U.S. Food and Drug Administration