Ongoing FDAEvent 99400 (2 months ago)

GE Healthcare MAC VU360 Electrocardiographs

After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity.

Moderate risk

Using it may cause temporary or medically reversible health problems. Serious harm is unlikely.

FDA Class II recall · How risk is rated

What to do

What happened

Software issue issue with the MAC VU360 v3.0 system. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.

From the FDA notice.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Affected if its UDI-DI matches.

  1. GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

    UDI-DI
    • 00198953107263

Details

Company began the recall
Sold in

Worldwide distribution - US Nationwide and the countries of Belgium, Hong Kong, Netherlands, Sweden, UK.

Event ID
99400
Recall No.
Z-2876-2026
Company
GE Medical Systems Information Technologies IncMilwaukee, Wisconsin
Quantity recalled
163 units
Recall type
Voluntary: Firm initiated
  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. Listed on Recall Canary

  4. In FDA's weekly enforcement report

  5. Summary updated