Carescape Central Station
Carescape Central Station may not resume patient monitoring for at least 3 minutes after restart, delaying recognition of events.
Moderate risk
Using it may cause temporary or medically reversible health problems. Serious harm is unlikely.
FDA Class II recall · How risk is rated
What happened
GE Medical Systems Information Technologies Inc of Waukesha, Wisconsin is recalling the Carescape Central Station CSCS V3 MAS800 Desktop and CSCS V3 MAI800 Integrated. The device may not resume patient monitoring for at least 3 minutes following system restart, which can lead to potential delay in recognition of clinically significant events. The recall began July 30, 2026, and was added to the FDA's Enforcement Report on September 11, 2026.
The recalled devices were distributed nationwide in the US and worldwide. The recall affects 29,570 units, with 23,632 in the US and 5,938 outside the US.
Summary by Recall Canary from the FDA notice.
What to do
We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:
- A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
- Talk to your health care provider before you stop using it.
Is yours affected?
Affected if its UDI-DI is listed.
Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV
- UDI-DI
- 00195278343512
- 00195278516367
- 00195278343383
Details
- Company began the recall
- Sold in
Worldwide - US Nationwide distribution.
- Event ID
- 99637
- Recall No.
- Z-3143-2026
- Company
- GE Medical Systems Information Technologies IncWaukesha, Wisconsin
- Hazard
- Device defect
- Quantity recalled
- 29,570 (23,632 US; 5,938 OUS)
- Recall type
- Voluntary: Firm initiated
- Customers told by
- Letter
Company began the recall
Added to FDA's Enforcement Report · Classified by FDA
Listed on Recall Canary
In FDA's weekly enforcement report
Summary updated
All recalls from GE Medical Systems Information Technologies Inc