FDA medical device recalls
Philips Respironics Trilogy EV300, Evo, Evo O2, and Evo Universal
In some situations, the obstruction alarm may not trigger within the timeframe dictated by relevant standards and may take up to four breaths.
4 productsCardinal Health Monoject Tuberculin Syringes
Packages labeled as Tuberculin Syringes contain U-100 Insulin syringes.
Medline Bronchoscopy and Airway Kits
Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves.
7 productseTRAX Needle Sensor and System Starter Kit
The eTRAX Needle Sensors and Needle System Starter Kits may identify the needle tip position incorrectly on the user interface.
8 productsBaxter Volara Ventilator Adapter Modules
The Blue Ventilator Adapter Module may cause oxygen desaturation or barotrauma in home care.
Abiomed Purge Cassettes and Impella Pump Sets
Purge Cassette has an increased risk of purge leaks.
Synthes 2.4 Volt Tap Locking and Cortex Screws
2.4 Volt TAP Locking Screw TAP to 90MM and 2.4 Volt(TM) TAP Cortex Screw to 90MM have an incorrect thread due to a lot swap.
2 productsOlympus SOLTIVE Premium and Pro Laser Systems
The Soltive laser systems may become inoperable due to a power supply defect, and smoke or a burning smell may occur.
2 productsDa Vinci S and Si Endoscopic Instruments
Da Vinci S and Si reusable instruments have broken or frayed grip cables.
4 productsDa Vinci S, Si Tenaculum Forceps and Permanent Cautery Hook Instruments
The da Vinci S, Si Tenaculum Forceps Instrument has frayed or broken pitch cables.
2 productsSumma Therapeutics Finesse BTK Multicath Injectable Angioplasty Balloon Catheter
The balloon in the device may not meet burst specifications.
Mentor Breast Tissue Expanders
Mentor Breast Tissue Expanders may have infusion sets with dull or blunt needle tips that may be difficult to advance and/or break.
7 productsPhilips Allura Xper and Azurion Angiography Systems
Allura Xper and Azurion systems may not initiate X-ray imaging or may do so intermittently when using the wired foot switch.
22 productsBD Kiestra ReadA
The BD Kiestra ReadA may have modules unreachable after a reboot, causing delays in plate retrieval.
Automated Impella Controller (AIC)
The Automated Impella Controller may delay its purge system blocked alarm when used with certain pumps.
Medline Enfit g-Tube Connector
The Enfit g-Tube Connector and Medline kits containing it may not seal properly, which could lead to leakage.
2 productsWaldemar Link Embrace Drill Towers
The Embrace Drill Tower has mixed-up article numbers in the surgical technique overview.
2 productsUltradent Jiffy Original Composite Polisher Cups
Jiffy Original Composite Polisher Cups may crumble and break apart easily, which could cause a delay in patient treatment.
RefleXion X1 Radiotherapy System
A software defect in the RefleXion X1 Radiotherapy System can cause the treatment plan dose to be delivered to the incorrect location.
Medline Sterile Radiology Procedure Kits
Sterile Radiology Procedure Kits may have a breach in sterile packaging, compromising sterility.
Stryker CHROMOPHARE Softlit Ring Surgical Light System
The surgical light assembly may not adequately support the weight of the ceiling cover.
ReCor Medical Paradise Ultrasound Renal Denervation Catheters
The Paradise Ultrasound Renal Denervation System catheter is nonconforming and not for human use.
2 productsPhilips Vue Motion V12
Vue Motion V12 may display image frames out of sequence during dynamic cine runs.
Puracol and Puracol Plus Collagen Wound Dressings
Puracol and Puracol Plus Collagen Wound Dressings have elevated endotoxin levels, which may cause a strong inflammatory response.
Atellica CH Enzymatic Hemoglobin A1c (A1c_E)
Atellica CH Enzymatic Hemoglobin A1c may give depressed results when used with another assay on the same analyzer.
Philips Respironics Trilogy Evo Ventilators
Using non-pneumatic nebulizers with these ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
4 productsMerit Medical Hemodialysis Catheters and Sheath Introducer
A design defect may cause the sheath introducer not to split as intended.
5 productsBD Alaris System with Guardrails Suite MX with Point of Care Unit
The labeling does not say which security version hospital networks must enable for secure Wi-Fi communication with Point-of-Care Units.
Olympus Cutting Forceps
Defective welds can result in the jaw breaking during clinical use.
7 productsTandem Pumps
Tandem pumps have a Czech user guide with translation errors that could lead to unintended insulin delivery.
OrthAlign Lantern Knee System Drill Plates
Drill plates are manufactured out of specification.
2 productsMedtronic MiniMed 780G Insulin Pumps
The software defects could result in a risk of hypoglycemia or hyperglycemia due to inaccuracies in insulin delivery.
Straumann Custom Abutments
A material mix-up occurred during manufacturing, and some abutments were made from TAN material instead of titanium.
2 productsTECNIS Eyhance IOL with TECNIS Simplicity Delivery System
The Tecnis Eyhance IOL may have a haptic that sticks to the optic, preventing the lens from unfolding consistently.
BiDop 3 Pocket Doppler Blood-Flow Detectors
The BiDop 3 Pocket Doppler has a fetal indication not cleared for use.
Philips Ultrasound Transducers
The notice provides clarification and labelling to define the useful life of ultrasound transducers in the field.
11 productsHandy Solutions Neck & Shoulders Heating Pad
The heating pad can cause excessive temperatures when folded or placed under the body, potentially causing burns.
I.T.S. Orthopedic Plates and Screws
I.T.S. Orthopedic Plates and Screws may cause higher RF-induced temperature increases under certain MRI conditions.
17 productsErbe Flexible Cryoprobes
Erbe is recalling its Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath and Flexible Cryoprobe because Probes may rupture/burst during activation.
3 productsOlympus Endoscope Valves and Reprocessors
The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
4 productsGE Medical Systems Imactis CT-Navigation Systems
The Imactis CT-Navigation System can lead to discordance between simulated and actual needle trajectory.
Beckman Coulter MicroScan Neg MIC 56 Panels
MicroScan Neg MIC 56 panels may be contaminated in some wells.
Datascope Battery Charging Station
The Battery Charging Station has a protruding screw that limits battery insertion and prevents proper charging.
Leksell Gamma Knife
Due to adapter not being properly aligned or locked in place.
Artegraft Vascular Graft
The Artegraft Vascular Graft has a labeling mix-up resulting in incorrect lot outer packaging.
ConvaTec EsteemBody Drainable Pouches
Convatec EsteemBody Drainable Pouch may leak due to a manufacturing issue.
3 productsAptima HPV Assay
The Aptima HPV Assay may generate invalid or false negative results.
Stryker Cub Pediatric Crib
The Cub Pediatric Crib is missing two access door warning labels.
STA Liatest D-Di
After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
Howmedica Osteonics MICS3 Angled Sagittal Saw Attachment
The MICS3 Angled Sagittal Saw Attachment has screws that may loosen and detach during use.