FDA medical device recalls
Lutronic DermaV Laser System
Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment.
HeartSine Samaritan PAD Defibrillators SAM 350P, SAM 360P and SAM 450P
HeartSine SAM 350P, SAM 360P, and SAM 450P defibrillators have a damaged membrane tail causing an intermittent fault.
Halyard IV Start Kits
Halyard IV Start Kit Non-Sterile lacks sterility assurance due to wrinkles in the paper lidding that may breach the seal area.
Bard Foley Catheter Trays and Intermittent Catheter Trays
Products that contain the Pericare Wipes may have contamination on the surface.
6 productsCardinal Health 200 Presource Packs Containing CliniClean Pericare Wipes Within BD Urinary Catheter Trays
Presource Packs with CliniClean Pericare Wipes in BD Urinary Catheter Trays may be contaminated with a lubricant, causing skin irritation.
12 productsFresenius Medical Care Acid Concentrates
Fresenius Medical Care Renal Therapies Group is recalling its Granuflo Dry Acid Concentrate because of complaints of blue-green discoloration in product.
3 productsFresenius Medical Care 5008X Blood Tubing Sets
Blood tubing sets may leak at the connection between the Arterial Alpha Clip and the main line, causing blood loss and hypovolemia.
4 productsB Braun Medical Discofix Four Way Stopcock Extension Sets
The Discofix Four Way Stopcock has an incorrect expiration date printed on the carton shipper label.
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System
The LUX-Dx Insertable Cardiac Monitor System may delay AF Burden Alert notification, potentially delaying clinical evaluation or therapeutic intervention.
3 productsMedos International Velys Enabling Tech Robotic-Assist Station
The Velys Enabling Tech Robotic-Assist Station has a software issue that may cause unintended movement of the robotic arm.
Philips Brilliance, IQon Spectral and Ingenuity CT Scanners
Philips CT systems have an unintended radiation issue that could affect equipment performance.
4 productsMedline Convenience Kits with BD ChloraPrep Applicators
Medline Convenience Kits containing BD ChloraPrep Clear Applicators may have an open or incomplete seal, which could lead to infection.
13 productsBaxter Novum IQ Syringe Pump
The syringe pump has a software issue that may occur during multi-syringe infusion.
2 productsJohnson & Johnson HEALON EndoCoat PRO and HEALON DUET PRO II
Viscoelastic products used in cataract procedures may cause corneal abrasion, iris trauma, or descemet's membrane detachment if forced.
Medline Convenience Kits
Medline Convenience Kits contain applicators with a potential breach of sterility, exposing patients and users to microbial contamination.
16 productsSTA Staclot Protein S
STA Staclot Protein S may show a positive drift in QC values within its stated 4-hour onboard stability period.
Halyard Medical Convenience Kits
Halyard medical convenience kits may contain neurological sponges with unexpected variability of endotoxin levels.
3 productsGE Portrait Core Services Software
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer.
4 productsICU Medical IV Sets with Drip Chambers
IV Sets with drip chambers may have black specks which could result in a delay in therapy.
Olympus POWERSEAL Curved Jaw Sealer & Divider
The outer shelf box is incorrectly labeled as a 37-cm device.
IMRIS Operating Room Tables ORT200 and ORT300
The operating room table may fail to lock, posing a risk of injury to the patient.
2 productsHealthmark ProChek Protein Tests
Healthmark ProChek-II and ProChek-W protein tests were made with an expired ingredient, which may cause false results.
2 productsFUJIFILM Synapse PACS Software
The software may cause the ruler measurement to be inaccurate.
Medline Convenience Kits
Medline convenience kits contain a potentially contaminated component.
11 productsCONMED Linvatec GENESYS CrossFT Suture Anchor with Disposable Driver
The suture anchor may contain an incorrect patient implant card, which may result in incorrect device identification and traceability.
Fresenius Medical Care 2008T BlueStar Hemodialysis Machine
Complaints were received about flow, no water and temperature alarms on 2008T BlueStar Hemodialysis machines.
Henry Schein Structur 2 SC and Grandio Restorative Materials
Specific lots were distributed in the US with incorrect packaging and Instructions for Use (IFU) from the U.S. FDA-cleared device.
2 productsMidmark Digital ECG HD+12 and IQecg Digital ECG
A communication problem may cause incorrect electrocardiogram waveforms and an inaccurate beats-per-minute calculation.
2 productsCanon Alphenix Angiography Systems
The Alphenix X-ray system has ceiling movement gear screws that may become loose.
Halyard Total Joint and DaVinci TruCustom Bundles
Halyard Total Joint and DaVinci TruCustom Bundles have labeling Deficiency.
7 productsMerge PACS and Merge OrthoPACS, Medical Imaging Diagnostic Workstation
Merge PACS and Merge OrthoPACS, a medical imaging diagnostic workstation, may display measurements inaccurately.
American Contract System Convenience Kits
The convenience kits contain catheters that may not meet performance specifications.
4 productsPhilips Incisive CT with Software V4.5&V5.0
Philips Incisive CT with software V4.5&V5.0 may fail to trigger a clinical scan during bolus tracking.
Medline Trach Tube Kit
The Medline medical convenience kit labeled as Trach Tube Kit has a component labeling discrepancy.
Hollister Onli Hydrophilic Urethral Intermittent Catheters
Onli 17cm Hydrophilic Urethral Intermittent Catheter may have a protrusion that could cause genitourinary trauma or bleeding.
Medline Medical Convenience Kits
The kits contain neurosponges with potential out-of-specification levels of endotoxin.
4 productsAmerican Contract Systems Medical Procedure Kits and Trays
Medical procedure kits and trays contain neurological sponges with potentially variable endotoxin levels.
2 productsBoston Scientific ENROUTE Transcarotid Neuroprotection System
Potential for arterial sheath tip separation or partial tip separation during use.
2 productsEsteem Body Soft Convex Urostomy Pouch
Esteem Body Soft Convex Urostomy Pouch may leak from cap during use.
BMC Luna G3 APAP Machines
Luna G3 Apap can shut down and display an error under extreme conditions due to a firmware defect.
Veritas Standard Glenoid Baseplate
The Veritas Standard Glenoid Baseplate has incorrectly machined peripheral screw holes, potentially causing improper screw engagement.
MiniMed 640G, 670G, and 780G Insulin Pump
Loaner insulin pumps may have been cleaned with incompatible products, damaging the case or keypad and causing insulin delivery issues.
3 productsFort Defiance Industries Automated Field Steam Sterilizer
The Automated Field Steam Sterilizer may have an intermittent loss of ground continuity in its water recovery system connector.
Siemens Healthcare Diagnostics Beckman Coulter and Syva EMIT 2000 Quinidine Assay
The assay may exhibit unacceptable precision performance, leading to erroneously elevated quinidine measurements.
2 productsBeckman Coulter DxC 500I Clinical Analyzer Modules
A software defect in the Parameter Package import logic overwrites predefined calculated test rules.
2 productsSpectral Instruments Lago X and Ami HTX Imaging Systems
Spectral Instruments is recalling its Lago X and Ami HTX because of Momentary absence of On-Screen "X-Ray On" banner during run readiness check.
2 productsParagon 28 Navicular Cage Foot Implants
The foot implant has elevated oxygen levels that may reduce material ductility.
2 productsBlanketrol II, Blanketrol III, and Innercool Coolblue Console
Overheating on 115V CPU boards may cause the device to shut down.
3 productsParagonix Technologies KidneyVault Portable Renal Perfusion System
The KidneyVault Portable Renal Perfusion System may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.
Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen
The Syphilis Screen assay may report low reactive results as nonreactive.
2 products