FDA•MODERATE
Instinct Plus Endoscopic Clipping Device
Wilson-Cook Medical is recalling the Instinct Plus Endoscopic Clipping Device due to a potential for the device to malfunction.

Wilson-Cook Medical is recalling the Instinct Plus Endoscopic Clipping Device due to a potential for the device to malfunction.

HydroPeptide Clear Alliance Serum recalled due to benzene contamination.
4 products
Cipla recalls Cinacalcet HCl 30 mg tablets over N-Nitroso impurity above acceptable limit.
3 products
Exchange Select Honey Lemon Cough Drops recalled over quality concerns from FDA inspection.
15 products
Remel is recalling Yeastone Broth due to the potential for incorrect antimicrobial susceptibility testing results.

LinkBio CORE Workstation (87-9136) may display incorrect Planning Date on Start Case menu.
Philips is recalling AneurysmFlow software because the MAFA ratio provides unreliable prognostic data for aneurysm treatment decisions.

Arjo Tenor patient lifts (Models KHA1000 US, KHA1010 US) recalled for actuator failure risk causing sudden arm drop.

Cipla recalls Lanreotide Injection 120 mg/0.5 mL over sterility assurance concerns.
2 products
GE Healthcare recalls Omnipaque (iohexol) Injection 350 mg/mL due to particulate matter.
2 products
Blaine Labs recalls multiple topical antiseptic/antifungal gels and creams due to Lysinibacillus fusiformis contamination.
9 products
Viatris is recalling Xanax XR 3 mg tablets (NDC 58151-506-91) due to failed dissolution specifications.

Fagron recalls Vancomycin HCl 1.5g/250mL injection due to sterility concern.
5 products
Geri-Care Artificial Tears Lubricant Eye Drops (0.5 oz, NDC 68788-7266-0) recalled due to lack of sterility assurance.

Medline recalls convenience kits with Namic RA syringes over risk of loose connection or disconnection during use.
25 products
Siemens is recalling ARTIS pheno and ARTIS Icono X-ray systems (Models 10849000, 11327600, 11328100, 11327700) due to erroneous dose display during patient exams.

Spacelabs is recalling Model 91496 Multi-parameter Command Module due to circuit board issue causing inaccurate cardiac output readings.

Molded Products is recalling See Luer Cap Set MPC-130 due to threaded sleeve detachment risk.

Fresenius Kabi recalls 14 lots of Sodium Chloride and Dextrose Injections due to sterility concerns.
14 products
Tornier is recalling shoulder system components incorrectly labeled as HRS Max parts, as they are incompatible with the HRS Max system.

Medline is recalling Anterior Hip Packs containing Olympus biopsy valves over risk of rubber fragment detachment.
8 products
GE Healthcare is recalling SIGNA Premier MR scanners due to ferrous magnet fittings that could cause personnel injury during service.

Medtronic MiniMed recalls all Paradigm, 600, and BLE 700 series insulin pumps over insulin delivery errors that can cause severe hypoglycemia or ketoacidosis.
14 products
Burlington Medical wrap aprons may lose shielding effectiveness over time.
13 products
Oxoid is recalling Microbact 12L Listeria Kit (MB1128A, Lot 4494873) due to risk of incorrect organism identification from ambiguous color reactions.
Philips Respironics recalls Trilogy Evo O2 ventilators (SW 1.05.15.00) over delayed obstruction alarm.
4 products
Cardinal Health is recalling Monoject Tuberculin Syringes (Catalog 1180100777) that contain U-100 insulin syringes instead, risking dosing errors.

American Health Packaging is recalling Meclizine Hydrochloride 12.5 mg tablets (NDC 60687-775-65 and 60687-775-11) due to failed tablet specifications.

Merit Medical recalls CentrosFLO Hemodialysis Catheters over sheath introducer defect that may cause hemorrhage and other serious complications.
5 products
Appco Pharma recalls Prazosin HCl 1mg capsules over nitrosamine impurity (NNP) above acceptable limits.
3 products
K.C. Pharmaceuticals recalls multiple brands of sterile eye drops over sterility concerns.
15 products
Amneal is recalling 500-count bottles of Tramadol Hydrochloride 50 mg tablets due to an impurity found during stability testing.

Chiesi USA is recalling Curosurf 240 mg intratracheal suspension vials (NDC 10122-510-03) due to lack of sterility assurance.

Synthes recalls VOLT Locking and Cortical Screw Taps due to incorrect thread from a lot swap.
2 products
Olympus recalls SOLTIVE laser systems (TFL-SLS, TFL-PLS) due to power supply defect that may cause inoperability and smoke.
2 products
Teva recalls Octreotide Acetate 20 mg vials over sterility assurance concerns.
6 products
A recall was issued for Artoura Breast Tissue Expanders due to infusion sets that may contain dull or blunt needle tips.
7 products
Philips recalls Allura Xper FD10/FD20 systems over wired foot switch failure causing intermittent X-ray imaging.
22 products
Medline recalls ENFit G-tube connectors in care/discharge kits over seal failure and leakage risk.
2 products
Medline is recalling Sterile Radiology Procedure Kits Model DYNDH1491B due to potential sterile pouch packaging breach.

Stryker is recalling the CHROMOPHARE Softlit Ring Surgical Light System due to a potential ceiling cover support failure.

ReCor Medical is recalling the Paradise Ultrasound Renal Denervation Catheter (Model PRDS-068-02) as these units were intended to be scrapped.
3 products
Philips Vue Motion V12 (1017979) may display image frames out of sequence during dynamic imaging runs.

Medline is recalling Puracol and Puracol Plus wound dressings due to elevated endotoxin levels that may cause severe inflammatory responses.

Siemens is recalling the Atellica CH A1c_E assay due to a risk of depressed results when used alongside the RCRP assay on the same analyzer.
Philips Respironics recalls Trilogy Evo ventilators over tidal volume discrepancy with non-pneumatic nebulizers.
4 products
BD Alaris System labeling lacks required TLS version specification for secure Wi-Fi communication with Point-of-Care Units.

Apotex is recalling Brimonidine Tartrate/Timolol Maleate eye drops (NDC 60505-0589-1) due to lack of sterility assurance.

Olympus recalls PKS Cutting Forceps over jaw breakage risk from defective welds.
7 products
Howmedica Osteonics is recalling the MICS3 Angled Sagittal Saw Attachment due to a risk of external screws loosening and detaching during use.