Medical device recalls
Paragon 28 Navicular Cage Foot Implants
The foot implant has elevated oxygen levels that may reduce material ductility.
2 productsBlanketrol II, Blanketrol III, and Innercool Coolblue Console
The Blanketrol II thermal regulating system may shut down due to overheating of its 115V CPU board.
3 productsParagonix Technologies KidneyVault Portable Renal Perfusion System
The KidneyVault Portable Renal Perfusion System may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.
Getinge Y Sensor Module, Servo-U, Servo-N
The module did not undergo electrical safety tests and functional tests during production.
Boston Scientific OverStitch Suture Cinch and Suture Cinch Long
The OverStitch Suture Cinch may fail to deploy the cinch implant or cut the suture.
2 productsTobii Dynavox TD Navio Maxi and Midi Speech Generating Device
The TD Navio Midi Speech Generating Device may have damaged battery mounting tabs, causing internal short circuit and heating.
2 productsKico Knee Innovation Arvis Surgical Navigation System GEN 2
The ARVIS surgical navigation system may show a tilted image due to an internal sensor issue.
Philips Cardiac Workstation 5000 and 7000
The Cardiac Workstation 5000 may provide inaccurate heart rate data.
2 productsDentsply Sirona Capsule Mixer
The Dentsply Sirona Capsule Mixer can discharge electricity when unplugged.
EndoFLIP System
The EndoFLIP System can incorrectly flag catheters as expired, making them unavailable for clinical procedures.
Cardinal Health Presource Kits
Procedure kits contain neurological sponges with unexpected variability in endotoxin levels.
8 productsBD Intravascular Administration and Extension Sets
Intravascular administration sets may not have received adequate sterilization processing.
Erbe USA ERBECRYO 2 Cryosurgical Unit
The ERBECRYO 2 Cryosurgical Unit can rupture its cryoprobe due to excessive pressure from a strong pull on the hose.
Straumann Bone Level Tapered (BLT) Implant
The Straumann Bone Level Tapered Implant may contain implants that are 12mm instead of 14mm.
Baxter Duo-Vent Solution Set and Male Luer Lock Adapter
The Baxter Duo-Vent Solution Set and Male Luer Lock Adapter can have air bubbles in the drip chamber and tubing.
Spectra Medical Devices Sodium Chloride Injection
Sodium Chloride Injection, 0.9%, USP 10mL Ampule may have compromised sterility, exposing patients to microorganisms, endotoxins, or foreign matter.
OBP Medical ER-Spec Single-Use Vaginal Speculum with LED Light Source
The ER-Spec vaginal speculum may have a burr on the blade, introducing a sharp edge for patients and providers.
Straumann GM Helix Dental Implants
There may be a mismatch between the implant contained inside the packaging and the information described on the outer carton label.
Philips Allura Xper FD Interventional Fluoroscopic X-Ray Systems
Allura Xper FD has a firmware issue that could lead to loss of geometry movements.
Terumo Glidesheath Slender A-Kit
The Glidesheath Slender was packaged with the 0.025 guidewire instead of the 0.021 guidewire.
Thermo Fisher Diagnostics Neisseria Meningitidis Gp C Tests
Neisseria meningitidis Gp C 2ml/VL is not performing to IFU criteria, not positive to ATCC13102 within 60 seconds.
Medline Micro-Kill Bleach Germicidal Bleach Wipes
Medline Micro-Kill Bleach Germicidal Bleach Wipes may have reduced bleach concentration before the end of their shelf life.
4 productsCovidien V-Loc, Dermalon and Surgipro Wound Closure Devices
V-Loc 90 Absorbable Wound Closure Device may have a perforation or tear in its sterile barrier packaging, which may lead to infection or prolonged surgery.
5 productsXhibit Central Station
The Xhibit Central Station has audio/video failures due to a display port component.
Roche Navify Digital Pathology Software
Navify Digital Pathology software has an image/metadata discrepancy with the Viewer window.
BOBBY Balloon Guide Catheter
The BOBBY Balloon Guide Catheter has an increased rate of balloon pinhole defects.
Contego Medical Neuroguard IEP Carotid Stent and Balloon System
The Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System may fail to open its integrated filter.
TISSIUM COAPTIUM Connect Peripheral Nerve Repair Devices
Currently labeled shelf life of 36 months is not substantiated by real-time data.
8 productsInstrumentation Laboratory GEM Premier 7000 and GEM Premier 5000 Blood Analyzers
The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 and GEM Premier 5000 PAK lots did not meet established manufacturing specifications.
2 productsMedline Hudson RCI Heated Neonatal Breathing Circuits
Neonatal and infant heated wire breathing circuits may experience sporadic failure.
3 productsSiemens Healthcare Diagnostics Atellica CH Urine Albumin
Atellica CH Urine Albumin tests may cause falsely depressed urea nitrogen and hemoglobin A1c results.
Beckman Coulter MicroScan LabPro Panel Update Software
MicroScan LabPro V5.03 Panel Update 11 has incorrect algorithms that may affect antimicrobial susceptibility test results.
2 productsPhilips Allura and Azurion X-Ray Systems
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
3 productsHardyDisk AST Cefiderocol 30µg (FDC30)
Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolonged infection, or worsening clinical outcome.
FilmArray NGDS Warrior Panel
The FilmArray NGDS Warrior Panel has mismatches between the outbreak strain and the reference strain used in its development.
GE Healthcare MAC VU360 Electrocardiographs
After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity.
Merge OrthoPACS and PACS
Merge PACS may show thumbnails from a previous patient study while another is open.
2 productsSyva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls
Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls have out-of-specification sodium azide concentration, causing low health risk.
Syva Emit Disopyramide Assay Enzyme Immunoassay
The Syva Emit Disopyramide Assay has an out-of-specification sodium azide concentration, which can cause mild skin or eye irritation.
Happiest Baby Pre-Loved SNOO Smart Sleeper
The pre-loved SNOO Smart Sleeper failed to meet quality standards before distribution.
Maquet Flow Anesthesia Machines
Events observed where anesthetic agent was not delivered when the vaporizer was activated.
Stryker Ahto Tube Sets and Suction/Irrigator Packaging
Stryker tube sets can leak from the handpiece during use.
5 productsMedical Action Industries Dialysis Change Tray Clear Sequence
The Dialysis Change Tray Clear Sequence has an improper seal that may cause doming or tearing of the silicone seal.
Innovasis Vector Pedicle Screw System Anti-Torque
The anti-torque instrument may have an undersized condition that could prevent it from engaging with the screw, leading to surgical delay.
Munevo DRIVE and Munevo DRIVE R-Net Powered Wheelchair Hands-Free Control Application
A software issue may cause the wheelchair's driving mechanism to fall out of calibration, leading to unexpected steering drift or imprecise control.
Boston Scientific Radial Jaw 4 Biopsy Forceps
Radial Jaw 4 Standard Capacity Biopsy Forceps may have a breach in the sterile barrier of their packaging.
3 productsFORWARD Oral Airway Kit and Kingfisher Compact Individual First Aid Kit
The FORWARD Oral Airway Kit may have an occluded or partially occluded Guedel Airway tip.
2 productsQuantum Rehab Power Wheelchairs
Quantum Rehab powered wheelchairs may show reminder pop-ups while in use, which can cause an unexpected park brake stop and lead to a fall or injury.
13 productsCharger for the Therabody Theragun, G4 PRO Massage Device
The charger for the Theragun G4 PRO may overheat, posing a fire risk.
Bard Access Systems BD Needle Kit for Powered Driver
BD Needle Kit for Powered Driver may have difficulty removing the obturator, potentially delaying therapy and causing serious injury or death.